BCI 1884SS Betriebsanleitung Seite 21

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 43
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 20
Chapter 4: Operating Instructions
Pulse Oximeter for Sleep Screening Operation Manual 4-7
Verification Mode
Verification Mode is intended for use in spot checking and for verifying that the user has
correct placement of the finger sensor.
The monitor will enter the Verification Mode as soon as trend memory is cleared.
In this mode the patient will be able to position the sensor on his/her finger (or appropriate
site) and obtain SpO
2
and heart rate data on the display.
This mode will last for 10 minutes of continuous valid data collection.
If the monitor is powered down during this mode, it will be powered up in this mode and
the verification timer will be reset.
Once the verification timer reads 10 minutes, the monitor exits theVerification Mode and
enters the Data Collection Mode.
The user functions in Verification Mode are:
The display will show % SpO
2
and heart rate values (if available, otherwise “dashes”
[ - - ] for invalid data are displayed).
The
x
key will turn the unit OFF without changing modes.
Pressing the F key momentarily will not activate the printer. The printer function is
suppressed in this mode.
Pressing and holding the
F key will advance the paper (see pg. 7-5) Reloading the
paper is allowed in this mode.
Pressing and holding the
F key at power up will erase trends, even if the trend
memory is already clear.
The serial link to the PC computer will be suppressed in this mode.
Data Collection Mode
Data Collection Mode is intended for use in long-term collection of data in sleep screening
environments.
The Data Collection Mode will be entered when the verification timer reaches 10 minutes.
Once the trend data collection process has been started, trend memory will be filled
continuously until the end-of-memory is detected (approx. 12 hours) or the monitor is
powered down.
Invalid data (such as sensor off) is stored in trend memory, but is not used in the statistical
calculations.
The user functions in Data Collection Mode are:
The SpO
2
display will show one flashing bar and heart rate display will be blank (see
Fig. 4.4).
Pressing and holding the
x
key for six (6) seconds will turn off the unit. While
pressing the
x
key, the flashing bar will flash at a higher frequency indicating that
the unit is about to be powered down.
NOTE! If the
x
key is released before the six seconds elapses, the bar will return to its
normal flashing pattern.
After the power down, the unit will power up in the Trend Printing Mode.